GeneProof Herpes Simplex Virus 2 (HSV-2) PCR Kit
Kat.: HSV2/ISEX/025
A termék CE/IVD minősítéssel rendelkezik és az OGYÉI által nyilvántartásba vett.
Az 2-es típusú Herpes Simplex (HSV-2) virus detektálására alkalmas master mix. A glycoprotein B-t (gB) kódoló gén specifikus, egyetlen kópiájú, konzervatív DNS szekvenciát célozza. (25 reakció)
Leírás
PRODUCT ADVANTAGES
UNIVERSAL PCR PROFILE
- Easily combinable with other PCR kits from the Immunocompromised / Transplanted, Neuroinfections and Sexually Transmitted Infections groups
- Simplifies laboratory workflow
CONTAMINATION ELIMINATION
- Ready-to-Use Master Mix contains Uracil-DNA glycosylase (UNG) and dUTPs eliminating possible contamination with amplification products
QUALITATIVE AND QUANTITATIVE DETECTION
- Quantification of viral load provides additional information for management of HSV infection
COMPATIBLE WITH A WIDE RANGE OF REAL-TIME PCR DEVICES
CE IVD CERTIFIED DIAGNOSTIC TEST
Technical specification
Indication | in vitro diagnostic medical device |
Intended User | For professional use in laboratories with trained staff |
Technology | real-time PCR |
Type of Analysis | qualitative and quantitative |
Target Sequence | specific conservative DNA sequence of a single-copy encoding the glycoprotein B (gB) |
Specificity | Herpes simplex virus type 2 (HSV-2), 100 % |
Sensitivity (LoD) | reaches up to 197,215 cp/ml with the probability of 95 % (on AcroMetrixTMHSV-2 Plasma Panel using manual extraction GeneProof PathogenFree DNA Isolation Kit) |
Linear Range | 1010 – 102.5 cp/ml with precision of ± 0.5 log |
Reporting Units | cp/μl |
Metrological Traceability | AcroMetrixTM HSV-2 Plasma Panel (AcroMetrix HSV-2) |
Extraction/Inhibition Control | PCR inhibition and DNA extraction efficiency control |
Validated Specimen | whole blood, plasma, urine, CSF |
Storage | -20 ± 5 °C |
Validated Extraction Method | croBEE NA16 Nucleic Acid Extraction System GeneProof PathogenFree DNA Isolation Kit |
Validated Instruments | croBEE Real-Time PCR System Applied Biosystems 7300 / 7500 Real-Time PCR System AriaMx Real-Time PCR System CFX Connect™ / CFX96™ / Dx Real-Time PCR Detection System LightCycler® 2.0 / 480 LineGene 9600 Plus Mic qPCR Cycler QuantStudio™ 3 / 5 Real-Time PCR System Rotor-Gene 3000 / 6000 / Q SLAN® Real-Time PCR System StepOne™ / StepOnePlus™ Real-Time PCR System |
Required Detection Channels | FAM, HEX |
External Quality Assessment | regularly tested by QCMD and Instand e.V. External Quality Assessment panels – for results see www.geneproof.com |
Certification | CE IVD for in vitro Diagnostics Use |