UVISDrop mikrotérfogatos spektrofotométer
Kat.: UVS-99
Hullámhossztartomány: 200-850 nm; a méréshez szükséges mintatérfogat: 0,5-2,5 ul; Fényforrás: Xenon lámpa
Leírás
UVISDropTM gives highly accurate full spectrum measurements with micro-samples (spectrum range: 200~ 850 nm). Simply pipette 0.5~ 2.5ul sample onto the pedestal, laying down the sampling arm, UVISDrop can automatically complete the measurement within 5 second
- UV/Vis measurement. Wide application, including nucleic acid (DNA/ RNA), protein, the concentration of other samples and counting of cell numbers
- A full spectrum measurement (200~ 850 nm) is available. The results of two light paths can be displayed simultaneously.
- No need to dilute high-concentration samples.
- No cuvettes or capillaries or other positioning instruments required.
- Buttons of “Blank” & “Measurement” on the shell, no need to use mouse.
- The light intensity is adjustable according to the sample concentrations to obtain stable measurements.
- Controllable by the computer software with data storage function.
Standard path length | 1 mm |
Short path length | 0.2 mm |
(for high concentration sample) | |
Sample volume | 0.5~ 2.5 ul |
Light source | Xenon flash lamp |
Detector | 3648-element linear silicon CCD array detector |
Wavelength range | 200 ~ 850 nm |
Wavelength accuracy | 1 nm |
Wavelength resolution | 3 nm (FWHM at Hg 546 nm) |
<1.8 nm (FWHM at Hg 253.7 nm) | |
Absorbance precision | 0.002 (1 mm path) |
Absorbance accuracy | 2% (at 0.76 absorbance at 257 nm) |
Measurement range | 0.1 ~ 70 Abs. |
Measurement accuracy | 0.1 ~ 5 Abs. ± 0.1 |
5 ~ 75 Abs. ± 2% | |
Measurement cycle | <5 seconds |
Size | 214 x147 x 95 mm |
Optical fiber | Quartz |
Voltage | 12 VDC |
Power | 6W |
Detects nucleic acid up to | 2~3500 ng/ul |
Detects protein up to | 100 mg/ml |
Weight | 1.4 kg |